低场磁共振技术与应用专栏

核磁共振技术测量生物骨密度的设备及方法

  • 李成伟 ,
  • 范东伟 ,
  • 姜佳岐 ,
  • 杜国红 ,
  • 王为民
展开
  • 1. 北京大学 信息科学技术学院量子电子学研究所, 北京 100871;
    2. 北京大学首钢医院, 北京 100144;
    3. 北京大学第三医院, 北京 100191
李成伟(1990-),男,江苏泰州人,硕士研究生,无线电物理专业.

收稿日期: 2016-03-15

  修回日期: 2016-07-11

  网络出版日期: 2016-09-05

基金资助

国家重点基础研究发展计划(“973计划”)资助项目(2015CB250902),国家自然科学基金资助项目(61531002).

Measuring Bone Mineral Density with NMR

  • LI Cheng-wei ,
  • FAN Dong-wei ,
  • JIANG Jia-qi ,
  • DU Guo-hong ,
  • WANG Wei-min
Expand
  • 1. Institute of Quantum Electronics, School of Electronics Engineering and Computer Science, Peking University, Beijing 100871, China;
    2. Shougang Hospital, Peking University, Beijing 100144, China;
    3. Peking University Third Hospital, Beijing 100191, China

Received date: 2016-03-15

  Revised date: 2016-07-11

  Online published: 2016-09-05

摘要

诊断骨质疏松症多采用双能X射线吸收方法测得的骨密度数据作为依据,但是该方法存在一定量的辐射,且准确性不高. 该文设计并改进了一套核磁共振装置,通过在发射和接收线圈之间增加快速切换开关(Q-switch),使得核磁共振自旋回波的回波时间缩短至80 ms的超短回波时间范围,实现了对T2弛豫时间很短的骨组织核磁共振自旋回波信号的采集;将T2弛豫时间谱和双能X射线方法测量的骨密度数据进行了对比研究,发现核磁共振T2弛豫时间谱上300~500 ms之间的峰值与骨密度之间具有良好的线性关系,随着骨密度的增大,相应的T2弛豫时间峰值缩短;利用核磁共振T2弛豫时间谱技术,结合双能X射线吸收方法,提出了一种测量生物离体骨骼样本骨密度的方法. 该测量方案具有测量速度快、对人体完全无害、跨时期比较性强等优点,很好的弥补了现有测量方法的不足,在骨质疏松症的诊断和疗效评估中有着良好的应用前景.

本文引用格式

李成伟 , 范东伟 , 姜佳岐 , 杜国红 , 王为民 . 核磁共振技术测量生物骨密度的设备及方法[J]. 波谱学杂志, 2016 , 33(3) : 468 -478 . DOI: 10.11938/cjmr20160312

Abstract

Bone mineral density (BMD) measured by dual energy X-ray absorptiometry is often used as an index for diagnosis of osteoporosis. However, ionization radiation is present in X-ray BMD measurement, and the result can be inaccurate in some caese. In this work, we constructed an NMR equipment that used a Q-switch to achieve echo time as low as 80 ms, and thus have the potential to be used for BMD measurement. In the T2 relaxation time spectra of bone measured by this equipment, the peak area at 300~500 ms was found to change linearly with the BMD measured by X-ray absorptiometry. Based on this observation, we put forward a method to measure in vitro BMD with NMR T2 relaxation time spectra. This kind of method has advantages such as high measuring speed, no ionization radiation, and high cross-period comparability.

参考文献

[1] Fanucci E, Manenti G, Masala S, et al.Multiparametercharacterisation of vertebral osteoporosis with 3-T MR[J]. Radiol Med, 2007, 112(2): 208-223.

[2] Techawiboonwong A, Song H K, Wehrli F W. In vivo MRI of submillisecond T2 species with two-dimensional and three-dimensional radial sequences and applications to the measurement of cortical bone water[J]. NMR Biomed, 2008, 21(1): 59-70.

[3] Techawiboonwong A, Song H K, Leonard M B, et al. Cortical bone water: in vivo quantification with ultrashort echo-time MR imaging[J]. Radiology, 2008, 248(3): 824-833.

[4] Du J, Carl M, Bydder M, et al. Qualitative and quantitative ultrashort echo time (UTE) imaging of cortical bone[J]. J Magn Reson, 2010, 207(2): 304-311.

[5] Wu Y, Ackerman J L, Chesler D A, et al. Density of organic matrix of native mineralized bone measured by water- and fat-suppressed proton projection MRI[J]. Magn Reson Med, 2003, 50(1): 59-68.

[6] Cao H, Ackerman J L, Hrovat M I, et al. Quantitative bone matrix density measurement by water- and fat-suppressed proton projection MRI (WASPI) with polymer calibration phantoms[J]. Magn Reson Med, 2008, 60(6): 1 433-1 443.

[7] Ni Q, Nyman J S, Wang X, et al. Assessment of water distribution changes in human cortical bone by nuclear magnetic resonance[J]. Meas Sci Technol, 2007, 18(3): 715-723.

[8] Nyman J S, Ni Q, Nicolella D P, Wang X. Measurements of mobile and bound water by nuclear magnetic resonance correlate with mechanical properties of bone[J]. Bone, 2008, 42(1): 193-199.

[9] Wang X, Ni Q. Determination of cortical bone porosity and pore size distribution using a low field pulsed NMR approach[J]. J Orthop Res, 2003, 21(2): 312-319.

[10] Fernandez S, Wehrli S L, Takahashi M, et al. Water content measured by proton-deuteron exchange NMR predicts bone mineral density and mechanical properties[J]. J Bone Miner Res, 2004, 19(2): 289-296.

[11] Horch R A, Nyman J S, Gochberg D F, et al. Characterization of 1H NMR signal in human cortical bone for magnetization resonance imaging[J]. Magn Reson Med, 2010, 64(3): 680-687.

[12] Reni B, Won B, Eric D, et al. Ultrashort echo time (UTE) imaging with bi-component analysis: Bound and free water evaluation of bovine cortical bone subject to sequential drying[J]. Bone, 2012, 50(3): 749-755.

[13] Won C B, Shantanu P, Reni B, et al. Magnetic resonance imaging assessed cortical porosity is highly correlated with μCTporosity[J]. Bone, 2014, 66: 56-61.

[14] Zhou Xiao-1ong(周小龙), Nie Sheng-dong(聂生东), Wang Yuan-jun(王远军), et al. A review on the inversion methods in 2D NMR(核磁共振二维谱反演技术综述)[J]. Chinese J Magn Reson(波谱学杂志), 2013, 30(2): 293-305.

[15] Wang Wei-min(王为民), Li Pei(李培), Ye Chao-hui(叶朝辉). Multi-exponential inversion method for NMR relaxation signals(核磁共振弛豫信号的多指数反演)[J]. Science in China(Series A)(中国科学A辑), 2001, 31(8): 730-736.

[16] Brownstein K R, Tarr C E. Importance of classic diffusion in NMR studies of water in biological cells[J]. Phy Rev A, 1979, 19(6): 2 446-2 453.

[17] Song Yi-qiao(宋一桥). Low field magnetic resonance: multi-dimensionalexperiments of relaxation and diffusion(低场核磁共振: 弛豫和扩散的多维实验)[J]. Chinese J Magn Reson(波谱学杂志), 2015, 32(2): 141-149.

[18] Joanne E H, Graeme M B. MR imaging with ultrashort TE (UTE) pulse sequences: Basic principles[J]. Radiography, 2005, 11(3): 163-174.

[19] William D L, Robert A A, Ghada E F, et al. Application of the 1994 WHO classification to populations other than postmenopausal Caucasian women: the 2005 ISCD official positions[J]. J Clin Densitom, 2006, 9(1): 22-30.

文章导航

/