该文提出了一种基于实际翻转角成像(Actual Flip-angle Imaging,AFI)的快速发射场测量方法(Fast AFI,FAFI),将多次激发平面回波成像(Multi-shot Echo Planar Imaging,Multi-shot EPI)的采集方式运用于AFI发射场(B1+)测量中,充分利用AFI序列中采集的等待时间,高倍数加速了水模和人体头部、腹部及盆腔的发射场测量.该文在水模和人体(n=16)实验中,验证了采用FAFI序列得到的B1+测量结果与AFI结果的一致性.FAFI序列大幅加速了发射场测量,为实现动态B1+匀场(B1+ shimming)和快速局部激发提供了高效的发射场测量方法.
张卫国
,
陈群
,
周子堃
,
胡凌志
,
蒋瑞瑞
,
贺强
. 基于AFI的快速发射场B1+成像方法[J]. 波谱学杂志, 2016
, 33(4)
: 515
-527
.
DOI: 10.11938/cjmr20160401
We propose a fast B1+ mapping method (FAFI) based on actual flip-angle imaging (AFI), in which the single-echo gradient echo (GRE) acquisition in the conventional AFI sequence is replaced by a multi-shot echo planar imaging (Multi-shot EPI) acquisition. FAFI makes full use of the waiting time, and thus has greatly increased imaging speed. We evaluated the performance of FAFI on phantoms and human subjects (n=16), and compared the results with those obtained using the conventional AFI method. It was demonstrated that it was possible to achieve high acceleration factor with FAFI, while preserving good consistency with the conventional AFI results. FAFI provides a powerful tool for dynamic B1+ mapping, and may find applications in dynamic B1+ shimming and rapid spatially selective radio frequency (RF) excitation.
[1] Alon L, Deniz C M, Brown R, et al. Method for in situ characterization of radiofrequency heating in parallel transmit MRI[J]. Magn Reson Med, 2013, 69(5):1457-1465.
[2] Suttie J J, Delabarre L, Pitcher A, et al. 7 Tesla (T) human cardiovascular magnetic resonance imaging using FLASH and SSFP to assess cardiac function:validation against 1.5 T and 3 T[J]. NMR Biomed, 2012, 25(1):27-34.
[3] Sprinkart A M, Kay N, Frank T, et al. Ultrafast volumetric B1+ mapping for improved radiofrequency shimming in 3 tesla body MRI[J]. J Magn Reson Imaging, 2014, 40(4):857-863.
[4] Collins C M, Zhangwei W. Calculation of radiofrequency electromagnetic fields and their effects in MRI of human subjects[J]. Magn Reson Med, 2011, 65(5):1470-1482.
[5] Homann H. SAR Prediction and Management for Parallel Transmit MRI[M]. German:KIT Scientific Publishing, 2012.
[6] Yang Bao-lian(杨保联). Future of ultra high field MRI in basic research and clinical applications(超高场磁共振人体成像应用研究和医学前景)[J]. Chinese J Magn Reson(波谱学杂志), 2015, 32(4):707-714.
[7] Collins C M, Wang Z, Mao W, et al. Array-optimized composite pulse for excellent whole-brain homogeneity in high-field MRI[J]. Magn Reson Med, 2007, 57(3):470-474.
[8] Son H W, Cho Y K, Gopinath A, et al. Shimming with SAR reduction in high-field MRI[J]. J Electromagnet Wavel, 2013, 27(12):1521-1524.
[9] Hernandez D, Min H C, Lee S Y. Iterative multi-channel radio frequency pulse calibration with improving B1 field uniformity in high field MRI[J]. Biomed Eng Online, 2015, 14(1):1-12.
[10] Katscher U, Börnert P. Parallel RF transmission in MRI[J]. NMR Biomed, 2006, 19(3):393-400.
[11] Cloos M A, Boulant N, Luong M, et al. Parallel-transmission-enabled magnetization-prepared rapid gradient-echo T1-weighted imaging of the human brain at 7 T[J]. Neuroimage, 2012, 62(3):2140-2150.
[12] Shao T, Xia L, Tao G, et al. Advanced three-dimensional tailored RF pulse design in volume selective parallel excitation[J]. IEEE Med Imaging, 2012, 31(5):997-1007.
[13] Insko E, Bolinger L. Mapping of the radio frequency field[J]. J Magn Reson Ser A, 1993, 103(1):82-85.
[14] Stollberger R, Wach P. Imaging of the active B1 field in vivo[J]. Magn Reson Med,1996, 35(1):246-251.
[15] Collins C M, Wanzhan Liu M S, Wang J, et al. Temperature and SAR calculations for a human head within volume and surface coils at 64 and 300 MHz[J]. J Magn Reson Imaging, 2004, 19(5):650-656.
[16] Yarnykh V L. Actual flip-angle imaging in the pulsed steady state:a method for rapid three-dimensional mapping of the transmitted radiofrequency field[J]. Magn Reson Med, 2007, 57(1):192-200.
[17] Bernstein M A, King K F, Zhou X J. Handbook of MRI Pulse Sequences[M]. USA:Elsevier Academic Press, 2004, 702.
[18] Nehrke K. On the steady-state properties of actual flip angle imaging (AFI)[J]. Magn Reson Med, 2009, 61(1):84-92.
[19] Buonocore M H, Gao L. Ghost artifact reduction for echo planar imaging using image phase correction[J]. Magn Reson Med, 1997, 38(1):89-100.
[20] Busse R F, Riederer S J. Steady-state preparation for spoiled gradient echo imaging[J]. Magn Reson Med, 2001, 45(4):653-661.
[21] Yuval Z. Two-dimensional phase correction method for single and multi-shot echo planar imaging[J]. Magn Reson Med, 2011, 66(6):1616-1626.
[22] Hsu Y C, Chern I, Zhao W, et al. Mitigate B1+ inhomogeneity using spatially selective radiofrequency excitation with generalized spatial encoding magnetic fields[J]. Magn Reson Med, 2014, 71(4):1458-1469.