Microwave Bridge in a DNP-EPR Multifunctional Spectrometer:Design and Implementation

  • HE Yu-gui ,
  • FENG Ji-wen ,
  • ZHU Tian-xiong ,
  • CHEN Jun-fei ,
  • LIU Chao-yang ,
  • LIU Zao
Expand
  • 1. State Key Laboratory of Magnetic Resonance and Atomic and Molecular Physics, National Center for Magnetic Resonance in Wuhan(Wuhan Institute of Physics and Mathematics, Chinese Academy of Sciences), Wuhan 430071, China;
    2. University of Chinese Academy of Sciences, Beijing 100049, China

Received date: 2016-04-19

  Revised date: 2017-07-14

  Online published: 2017-09-05

摘要

在自主研制的动态核极化(Dynamic Nuclear Polarization,DNP)分子影像装置的基础上,提出了一种集DNP和电子顺磁共振(Electron Paramagnetic Resonance,EPR)于一体的多功能谱仪,并对其中的关键部件之一——微波桥进行了设计.微波桥的引入,实现了DNP微波发射机的集成化,以及在DNP谱仪基础上的EPR功能扩展.通过结构设计、电路仿真及系统测试,完成了高频谱纯度、高动态范围的微波发射机以及低噪声系数的微波检测系统的设计与制作.并通过DNP增强实验以及连续波EPR实验对微波桥的性能进行了验证.

本文引用格式

贺玉贵 , 冯继文 , 朱天雄 , 陈俊飞 , 刘朝阳 , 刘早 . DNP-EPR多功能谱仪中微波桥的设计与实现[J]. 波谱学杂志, 2017 , 34(3) : 357 -364 . DOI: 10.11938/cjmr20162517

Abstract

A multi-functional spectrometer, capable of dynamic nuclear polarization (DNP) and electron paramagnetic resonance (EPR) experiments, is put forward on the basis of a home-built DNP-magnetic resonance imaging (MRI) spectrometer. A key component of the multi-functional spectrometer, microwave bridge, was designed and implemented. The microwave bridge was used to integrate the microwave transmitter of the DNP spectrometer and to enable the EPR capability. A transmitter system with high spectrum purity and dynamic range as well as a detector with low noise figure were also implemented by structure designing, circuit simulation and system evaluation. DNP and continuous wave EPR experiments were conducted to verify the reliability of the microwave bridge.

参考文献

[1] 裘祖文, 裴奉奎. 核磁共振波谱[M]. 北京:科学出版社, 1989
[2] OVERHAUSER A W. Polarization of nuclei in metals[J]. Phys Rev, 1953, 92:411-415.
[3] TAO Q, HE Y G, WANG C, et al. A TM110 dynamic nuclear polarization probe:design and application[J]. Chinese J Magn Reson, 2016, 33(1):44-53. 陶泉, 贺玉贵, 王超, 等. TM110模式DNP探头的研制与应用[J]. 波谱学杂志, 2016, 33(1):44-53.
[4] MALY T, DEBELOUCHINA G T, BAJAJ V S, et al. Dynamic nuclear polarization at high magnetic fields[J]. J Chem Phys, 2008, 128:052211.
[5] HE Y G, FENG J W, ZHANG Z, et al. Development of pulsed dynamic nuclear polarization for enhancing NMR and MRI[J]. Chinese J Magn Reson, 2015, 32(2):393-398. 贺玉贵, 冯继文, 张志, 等. 脉冲动态核极化增强的NMR和MRI系统研究[J]. 波谱学杂志, 2015, 32(2):393-398.
[6] 卢景雰. 现代电子顺磁共振波谱学及其应用[M]. 北京:北京大学出版社, 2012
[7] 荣星. 脉冲电子顺磁共振谱仪研制及应用[D]. 合肥:中国科学技术大学, 2011.
[8] KAUFMANN T, KELLER K J, FRANCK J M, et al. DAC-board based X-band EPR spectrometer with arbitrary waveform control[J]. J Magn Reson, 2013, 235:95-108.
[9] 张玉兴, 杨玉梅, 敬守钊. 射频模拟电路与系统[M]. 成都:电子科技大学出版社, 2008.
[10] LUDING R, BRETCHKO P著. 王子宇, 王心悦译. 射频电路设计:理论与应用[M]. 北京:电子工业出版社, 2002.
文章导航

/