以万里长城为代表的古建筑是世界瑰宝,更是中华民族的象征和骄傲.本文提出利用便携式核磁共振(NMR)装置来探测研究这类古建筑的建筑材料,在不对其造成损伤的基础上,发掘其隐含的科学、技术和工程相关的丰富信息.为此,作为第一步,设计了适合于探测这类古建筑的便携式单边NMR探测器组合式磁体.该探测器的磁体结构以semi-Halbach为基础,通过不同磁体模块间的组合得到对应移动探测模式、长距离探测模式和均匀磁场探测模式的磁体结构.随后根据优化结果,设计加工了磁体组件,并采用该磁体进行了流体、长城城砖和现代红砖的NMR实验,实测结果与模拟一致.该组合式磁体的优点在于通过不同磁体模块组合,实现了多种探测方式,适用于探测长城等这类古建筑物需要多种探测模式的科学研究.
唐炼
,
陈伟梁
,
赵昆
,
孙哲
,
周华
,
王杰
,
张岩
,
刘化冰
,
相文峰
,
肖立志
,
廖广志
. 用于长城等古建筑探测的NMR探测器的磁体设计[J]. 波谱学杂志, 2017
, 34(3)
: 372
-382
.
DOI: 10.11938/cjmr20172569
Ancient architectures such as the Great Wall are symbols and pride of the Chinese nation, and precious treasures of the world. Portable NMR detection devices have been proposed to examine the building materials of such ancient architectures non-invasively, and to provide information regarding the science, technology and engineering behind these architectures. In this study, the magnets for such portable NMR detection devices were designed and built, which were all based on a semi-Halbach magnet structure. On top of the basic design, different modules were added to achieve different functionality, including the magnet for mobile measurement, the magnet for high depth measurement, and the "homogeneous" field mode. The measured magnetic field (B0) distributions for these magnets agreed well with the simulated results, with only slight differences. Experiments on bricks were performed to verify the performance of these magnets.
[1] COOPER R K, JACKSON J A. Remote (inside-out) NMR. I. Remote production of a region of homogeneous magnetic field[J]. J Magn Reson, 1980, 41(3):400-405.
[2] JACKSON J A. Nuclear magnetic resonance well logging[J]. Log Analyst. 1984, 25(5):16-30
[3] CHANG W H, CHEN J H, HWANG L P. Single-sided mobile NMR with a Halbach magnet[J]. Magn Reson Imaging, 2006, 24(8):1095-1102.
[4] EIDMANN G, SAVELSBERG R, BLUMLER P, et al. The NMR MOUSE, a mobile universal surface explorer[J]. J Magn Reson, 1996, 122(1):104-109.
[5] PRADO P J. Single sided imaging sensor[J]. Magn Reson Imaging, 2003, 21(3,4):397-400.
[6] KLEINBERG R L, SEZGINER A, GRIFFIN D D, et al. Novel NMR apparatus for investigating an external sample[J]. J Magn Reson, 1992, 97(3):466-485.
[7] MARBLR A E, MASTIKHIN I V, COLPITTS B G, et al. A constant gradient unilateral magnet for near-surface MRI profiling[J]. J Magn Reson, 2006, 183(2):78-87.
[8] MANZ B, COY A, DYKSTRA R, et al. A mobile one-sided NMR sensor with a homogeneous magnetic field:the NMR-MOLE[J]. J Magn Reson, 2006, 183(1):25-31.
[9] MCDONALD P J, AKHMEROV A, BACKHOUSE L J, et al. Magnetic resonance profiling of human skin in vivo using GARField magnets[J]. J Pharm Sci, 2005, 94(8):1850-1860.
[10] PERLO J, CASANOVA F, BLUMICH B. Ex situ NMR in highly homogeneous fields:1H spectroscopy[J]. Science, 2007, 315(5815):1110-1112.
[11] DIAS M, HADGRAFT J, GLOVER P M, et al. Stray field magnetic resonance imaging:a preliminary study of skin hydration[J]. J Phys D Appl Phys, 2003, 36(4):364-368.
[12] BLUMICH B, PERLO J, CASANOVA F. Mobile single-sided NMR[J]. Prog Nucl Mag Res Sp, 2008, 52(4):197-269.
[13] ZALESSKIY S S, DANIELI E, BLUMICH B, et al. Miniaturization of NMR systems:Desktop spectrometers, microcoil spectroscopy, and "NMR on a chip" for chemistry, biochemistry, and industry[J]. Cheminform, 2014, 45(11):5641-5694.
[14] COATES G R, XIAO L Z, PRAMMER M. NMR Logging Principles and Applications[M]. USA:Gulf Professional Publishing, 1999.
[15] LIU F G, YANG Q X. Magnetic scaler potential finite element method applied in 3-D permanent magnetic fields[J] Journal of Hebei University of Technology, 1997, 3:49-54. 刘福贵, 杨庆新. 三维永久磁铁磁场的标量磁位数值计算[J]. 河北工业大学学报, 1997, 3:49-54.
[16] 金建铭. 电磁场有限元方法[M]. 西安:西安电子科技大学出版社, 1998
[17] HALBACH K. Strong rare earth cobalt quadrupoles[J]. IEEE T Nucl Sci, 1979, 26(3):3882-3884.
[18] SOLTNER H, BLUMER P. Dipolar Halbach magnet stacks made from identically shaped permanent magnets for magnetic resonance[J]. Concept Magn Reson A, 2010, 36A(4):211-222.
[19] WINDT C W, SOLTNER H, VAN D D, et al. A portable Halbach magnet that can be opened and closed without force:the NMR-CUFF[J]. J Magn Reson, 2011, 208(1):27-33.
[20] ANFEROVA S, ANFEROV V, ADAMS M, et al. Construction of a NMR-MOUSE with short dead time[J]. Concept Magn Res, 2002, 15(1):15-25.
[21] HE W, HE X L, XU Z, et al. The Gram-Schmidt data fitting method for the designing of the gradient magnetic field of unilateral NMR[J] Journal of Chongqing University (Natural Science Edition), 2013, 36(1):86-91. 何为, 何晓龙, 徐征, 等. 单边核磁共振磁体梯度磁场设计的Gram-Schmidt正交化拟合方法[J]. 重庆大学学报(自然科学版), 2013, 36(1):86-91.
[22] Institute for Applied Physics. Dielectric properties of body tissues[EB/OL].[2011-01-28]. http://niremf.ifac.cnr.it/tissprop/.
[23] CARR H Y, PURCELL E M. Effects of diffusion on free precession in nuclear magnetic resonance experiments[J]. Phys Rev, 1954, 94(3):630-638.
[24] MEIBOOM S, GILL D. Modified spin echo method for measuring nuclear relaxation times[J]. Review of Scientific Instruments, 1958, 29:668-691.
[25] CALLAGHAN P T. Principles of magnetic resonance microscopy[M]. Oxford:Clarendon Press, 1991.
[26] SUN Z, XIAO L Z, LIAO G Z, et al. A modular single-sided NMR sensor design with multifunction[C]. Bologna Italy:13th International Bologna Conferenceon Magnetic Resonance in Porous Media, 2016.
[27] SUN Z, XIAO L Z, ZHANG Y, et al. A modular and multi-functional single-sided NMR sensor[J]. Micropor Mesopor Mat, (in press). DOI:http://doi.org/10.1016/j.micromeso.2017.05.039.