设计了一种基于现场可编程门阵列(FPGA)的磁共振成像(MRI)谱仪B0信号的高精度发生方法,并对产生的B0信号经高速采集卡采集之后进行测试和验证.FPGA从外部读取波形数据和参数,分别存储在双端口随机存取存储器(RAM)和参数寄存器中,根据预补偿算法实现B0信号的发生,并通过对时间参数和幅度参数的控制,产生不同的B0信号,时间分辨率为1 μs.对谱仪的B0输出进行采集,再进行最小二乘拟合,以验证B0信号发生的准确性.经实验证实,该设计可以产生正确、可控的高精度B0信号.
A field programmable gate array (FPGA)-based method is proposed to generate B0 signals, with high accuracy, on magnetic resonance imaging (MRI) spectrometers. The FPGA reads the waveform data and external parameters, and stores them in the dual-port random access memory (RAM) and parameter registers, respectively. The B0 signals are then generated using the pre-emphasis algorithm, and different time or amplitude parameters give different B0 signals. The B0 signals are collected by high speed acquisition card, and the least squares fitting algorithm is used to test the accuracy of the B0 signals generated. The performance of this method was validated by experiments.
[1] LIU M, QIU W Q, Sun H J, et al. Progress in the portable NMR spectrometer[J]. Chinese J Magn Reson. 2014, 31(4):504-514. 刘敏, 邱雯琦, 孙慧军, 等. 便携式核磁共振谱仪的研究进展[J]. 波谱学杂志, 2014, 31(4):504-514.
[2] SHEN J, LIU Y, LI J Q, et al. A powerful graphical pulse sequence programming tool for magnetic resonance imaging[J]. MAGMA, 2005, 18(6):332-342.
[3] MAO W P, BAO Q J, YANG L, et al. A modularized pulse programmer for NMR spectroscopy[J]. Meas Sci Technol, 2011, 22(2):025901.
[4] STANG P P, CONOLLY S M, SANTOS J M, et al. Medusa:a scalable MR console using USB[J]. IEEE Trans Med Imaging, 2012, 31(2):370-379.
[5] HOU S L, XIE H T, HOU X W, et al. Gradient coil of permanent magnet mini-magnetic resonance imager and image quality[J]. Chinese J Magn Reson. 2014, 29(4):508-520. 侯淑莲, 谢寰彤, 侯晓吻, 等. 永磁微型磁共振成像仪的梯度线圈与图像质量[J]. 波谱学杂志, 2014, 29(4):508-520.
[6] XU Q, WANG H, JIANG Y, et al. A digital preemphasis gradient waveform generator for magnetic resonance imaging[J]. Chinese J Magn Reson, 2006, 23(1):11-16. 徐勤, 王鹤, 蒋瑜, 等. 一种具有数字预加重的磁共振成像梯度波形发生器[J]. 波谱学杂志, 2006, 23(1):11-16.
[7] XIAO L, TANG W N, WANG W M. An FPGA-based single-chip gradient calculation module for magnetic resonance imaging[J]. Chinese J Magn Reson, 2010, 27(2):163-171. 肖亮, 汤伟男, 王为民. 基于单片FPGA的磁共振成像梯度计算模块[J]. 波谱学杂志, 2010, 27(2):163-171.
[8] ZANG F C, LI J Q, WANG H, et al. Automatic adjustment of gradient preemphasis in MRI systems[J]. Chinese J Magn Reson, 2008, 25(1):26-32. 臧凤超, 李建奇, 王鹤, 等. 磁共振成像系统中的自动梯度预加重调节方法研究[J]. 波谱学杂志, 2008, 25(1):26-32.
[9] LIU H D, WANG H W, ZHU T X, et al. A frequency synthesizer for high-field NMR spectrometers[J]. Chinese J Magn Reson, 2014, 31(1):91-99. 刘海丹, 王辉旺, 朱天雄, 等. 高场核磁共振波谱仪频率合成器设计[J]. 波谱学杂志, 2014, 31(1):91-99.
[10] FRY M E, PITTARD S, SUMMERS I R, et al. A programmable eddy-current compensation system for MRI and localized spectroscopy[J]. J Magn Reson Imaging, 1997, 7(2):455-458.
[11] VARDAL J, SALO R A, LAESSON C, et al. Correction of B0-distortions in echo-planar-imaging-based perfusion-weighted MR[J]. J Magn Reson Imaging, 2014, 39(3):722-728.