研究论文

低场永磁体磁共振射频场映像

  • 陈海燕 ,
  • 赵世龙 ,
  • 李晓南 ,
  • 刘国强 ,
  • 胡丽丽 ,
  • 刘弢
展开
  • 1. 省部共建电工装备可靠性与智能化国家重点实验室(河北工业大学 电气工程学院), 天津 300130;
    2. 河北省电磁场与电器可靠性重点实验室(河北工业大学 电气工程学院), 天津 300130;
    3. 中国科学院 电工研究所, 北京 100190;
    4. 北京纳志医疗科技有限公司, 北京 100089

收稿日期: 2018-06-08

  网络出版日期: 2018-07-10

基金资助

国家自然科学基金资助项目(51577184);北京市自然科学基金资助项目(7182111);国家自然科学基金青年科学基金资助项目(51407174).

B1 Mapping on Low-Field Permanent Magnet MRI Scanner

  • CHEN Hai-yan ,
  • ZHAO Shi-long ,
  • LI Xiao-nan ,
  • LIU Guo-qiang ,
  • HU Li-li ,
  • LIU Tao
Expand
  • 1. State Key Laboratory of Reliability and Intelligence of Electrical Equipment, School of Electrical Engineering, Hebei University of Technology, Tianjin 300130, China;
    2. Key Laboratory of Electromagnetic Field and Electrical Apparatus Reliability of Hebei Province, School of Electrical Engineering, Hebei University of Technology, Tianjin 300130, China;
    3. Institute of Electrical Engineering, Chinese Academy of Sciences, Beijing 100190, China;
    4. Beijing Nazhi Health Care Co., Ltd., Beijing 100089, China

Received date: 2018-06-08

  Online published: 2018-07-10

摘要

射频场映像是通过一定算法对磁共振射频线圈的发射场进行重建的方法.高场下的射频场经过生物组织时会发生明显变化,在其基础上可以反演生物组织体内电特性,进而对癌症等疾病进行早期诊断,是对生物组织的磁共振结构成像的有力补充.目前为止,射频场映像和电特性研究都以高场鸟笼线圈为主,对低场下的相控阵研究较少.本文主要研究了低场永磁体磁共振射频场的均匀度.有限元仿真和实验验证了在17.8 MHz激励下,射频场在空载和负载下均匀度均发生较大变化.射频场均匀度在负载下的改变在一定程度上可以反映负载生物组织的电特性,对磁共振电特性实用化研究提供了一定的参考价值.

本文引用格式

陈海燕 , 赵世龙 , 李晓南 , 刘国强 , 胡丽丽 , 刘弢 . 低场永磁体磁共振射频场映像[J]. 波谱学杂志, 2018 , 35(4) : 498 -504 . DOI: 10.11938/cjmr20182661

Abstract

Methods to map radiofrequency (RF) field generated by an RF coil/pulse have been developed. By measuring the changes of RF field in biological tissues when an RF pulse passing through, the electrical properties of the tissue could be retrieved and used for early diagnosis of diseases such as cancer. So far, RF field mapping is mainly done for birdcage coils at high fields. Less research has been performed on phased array coils at low fields. This work studied how loading would affect the distribution of RF field in low-field MRI with permanent magnet (i.e., 17.8 MHz). Both finite element simulation and experimental results demonstrated that coil loading changed the uniformity of RF field significantly, and the distortion of RF field under the loaded condition could reflect the electrical characteristics of the loaded biological tissues.

参考文献

[1] JOINES W T, ZHANG Y, LI C X, et al. The measured electrical properties of normal and malignant human tissues from 50 to 900 MHz[J]. Med Phys, 1994, 21(4):547-550.
[2] LIU J E, ZHANG X T, VAN DE MOORTELE P F, et al. Determining electrical properties based on B1 fields measured in an MR scanner using a multi-channel transmit/receive coil:a general approach[J]. Phys Med Biol, 2013, 58(13):4395-4408.
[3] KATSCHER U, ABE H, IVANCEVIC M, et al. Systematic brain tumor conductivity study with optimized EPT sequence and reconstruction algorithm[C] 201321th Annual International Conference of the ISMRM. Salt Lake City:2013:33-40.
[4] HAEMMERICH D, STAELIN S T, TSAI J Z, et al. In vivol electrical conductivity of hepatic tumours[J]. Physiol Meas, 2003, 24(2):251-260.
[5] FRICKE H. The electric capacity of suspensions with special reference to blood[J]. J Gen Physiol, 1925, 9(2):137-152.
[6] HAACKE E M, PETROPOULOS L S, NILGES E W, et al. Xtraction of conductivity and permittivity using magnetic resonance imaging[J]. Phys Med Biol, 1991, 36(6):723-734.
[7] WEN H. Noninvasive quantitative mapping of conductivity and dielectric distributions using RF wave propagation effects in high field MRI[C]. Proc SPIE 5030, Medical Imaging 2003:Physics of Medical Imaging, 2003, doi:10.1117/12.480000.
[8] KATSCHER U, VOIGT T, FINDEKLEE C, et al. Determination of Electric conductivity and local SAR Via B1 mapping[J]. IEEE Trans Med Imaging, 2009, 28(9):1365-1374.
[9] ZHANG X T, SCHMITTER S, VAN DE MOORTELE P F, et al. From complex B1 mapping to local SAR estimation for human brain MR imaging using multi-channel transceiver coil at 7T[J]. IEEE Trans Med Imaging, 2013, 32(6):1058-1067.
[10] YANG B L. Future of ultra high field MRI in basic research and clinical applications[J]. Chinese J Magn Reson, 2015, 32(4):707-714. 杨保联. 超高场磁共振人体成像应用研究和医学前景[J]. 波谱学杂志, 2015, 32(4):707-714.
[11] 赵喜平. 磁共振成像[M]. 北京:科学出版社, 2004.
[12] ZHOU Z K, HU L Z, JIANG R R, et al. Rapid B1+ mapping based on actual flip-angle imaging[J]. Chinese J Magn Reson, 2016, 33(4):515-527. 周子堃, 胡凌志, 蒋瑞瑞, 等. 基于AFI的快速发射场B1+成像方法[J]. 波谱学杂志, 2016, 33(4):515-527.
[13] HE X W, YUAN R B, LI B K, et al. The design of an open MRI 4-channel receive-only phased array knee coil[J]. Appl Magn Reson, 2016, 47(5):499-510.
[14] XIN X G. Technological progresses of the magnetic resonance electrical property tomography of human tissues[J]. Chinese Journal of Biomedical Engineering, 2015, 34(1):83-90. 辛学刚. 人体组织电特性磁共振断层成像(MR EPT)技术进展[J]. 中国生物医学工程学报, 2015, 34(1):83-90.
文章导航

/