研究论文

基于法布里-珀罗谐振腔的W波段电子顺磁共振探头研制

  • 罗文尤 ,
  • 荣星
展开
  • 中国科学院微观磁共振重点实验室,中国科学技术大学,安徽 合肥 230026
* Tel: 0551-63606664, E-mail: xrong@ustc.edu.cn.

收稿日期: 2024-04-19

  网络出版日期: 2024-06-11

基金资助

国家自然科学基金资助项目(12150010)

A W-Band Electron Paramagnetic Resonance Probe Based on Fabry-Perot Cavity

  • LUO Wenyou ,
  • RONG Xing
Expand
  • CAS Key Laboratory of Microscale Magnetic Resonance and School of Physical Sciences, University of Science and Technology of China, Hefei 230026, China

Received date: 2024-04-19

  Online published: 2024-06-11

摘要

本文研制了一种基于法布里-珀罗谐振腔的W波段电子顺磁共振探头.利用法布里-珀罗谐振腔所具备的大体积和开放腔等特点,本文将法布里-珀罗谐振腔用作谱仪探头的谐振器部分,并且设计了波纹波导来实现微波桥与法布里-珀罗谐振腔之间的低损耗微波信号传输.所研制的探头工作在W波段,空载品质因数为1 542.最终将该探头集成于W波段电子顺磁共振谱仪,通过测量Mn(II)掺杂的CaO粉末样品的谱线,展示了该探头在常温下用于W波段连续波电子顺磁共振实验的能力.探头的调制场频率设置为100 kHz,实验结果显示探头的最大调制场幅度可达8.9 G(1 G=10-4 T),谱仪绝对自旋数灵敏度为6.6×108 spins/(G•$\sqrt{Hz}$).

本文引用格式

罗文尤 , 荣星 . 基于法布里-珀罗谐振腔的W波段电子顺磁共振探头研制[J]. 波谱学杂志, 2024 , 41(4) : 393 -404 . DOI: 10.11938/cjmr20243111

Abstract

This paper presents a new W-band electron paramagnetic resonance (EPR) probe design based on Fabry-Perot cavity. In this probe, a Fabry-Perot cavity with open structure and large effective volume is used as a resonator. Corrugated waveguides are designed to achieve low-loss microwave signal transmission between the W-band microwave bridge and the Fabry-Perot cavity. The probe works in W-band, and the unloaded quality factor is 1 542. A room-temperature continuous wave EPR experiment with Mn(II) in CaO powder was conducted after assembling the probe into a W-band EPR spectrometer. The results show that the modulation field amplitude is 8.9 G (1 G=10-4 T) when the modulation field frequency is set to 100 kHz, and the absolute spin sensitivity is estimated to be 6.6×108 spins/(G•$\sqrt{Hz}$), which verifies the feasibility of utilizing Fabry-Perot cavity as a high sensitivity W-band EPR probe.

参考文献

[1] M?BIUS K, SAVITSKY A, SCHNEGG A, et al. High-field EPR spectroscopy applied to biological systems: characterization of molecular switches for electron and ion transfer[J]. Phys Chem Chem Phys, 2005, 7(1): 19-42.
[2] DUSS O, YULIKOV M, JESCHKE G, et al. EPR-aided approach for solution structure determination of large RNAs or protein-RNA complexes[J]. Nat Commun, 2014, 5(1): 3669.
[3] ROESSLER M M, SALVADORI E. Principles and applications of EPR spectroscopy in the chemical sciences[J]. Chem Soc Rev, 2018, 47(8): 2534-2553.
[4] BABUNTS R A, GURIN A S, EDINACH E V, et al. Non-Kramers iron S = 2 ions in β-Ga2O3 crystals: high-frequency low-temperature EPR study[J]. J Appl Phys, 2022, 132(15): 155703.
[5] PENG Z, DALLAS J, TAKAHASHI S. Reduction of surface spin-induced electron spin relaxations in nanodiamonds[J]. J Appl Phys, 2020, 128(5): 054301.
[6] HE Y, SHI Z F, ZHAO X X, et al. Design and performance of a new multifunction X-band EPR spectrometer[J]. Chinese J Magn Reson, 2022, 39(1): 1-10.
  贺羽, 石致富, 赵新星, 等. 新型X波段多功能EPR谱仪的设计与性能[J]. 波谱学杂志, 2022, 39(1): 1-10.
[7] YU Y Y, LI X Y, SUN B, et al. Design and optimization of terahertz directional coupler based on hybrid-cladding hollow waveguide with low confinement loss[J]. Chin Phys B, 2015, 24(6): 068702.
[8] PARIZH M, LVOVSKY Y, SUMPTION M. Conductors for commercial MRI magnets beyond NbTi: requirements and challenges[J]. Supercond Sci Technol, 2016, 30(1): 014007.
[9] FEHER G. Sensitivity considerations in microwave paramagnetic resonance absorption techniques[J]. Bell Syst Tech J, 1957, 36(2): 449-484.
[10] M?BIUS K, SAVITSKY A. High-field/high-frequency EPR spectroscopy in protein research: principles and examples[J]. Appl Magn Reson, 2023, 54(2): 207-287.
[11] TELSER J, KRZYSTEK J, OZAROWSKI A. High-frequency and high-field electron paramagnetic resonance (HFEPR): a new spectroscopic tool for bioinorganic chemistry[J]. J Biol Inorg Chem, 2014, 19: 297-318.
[12] ROHRER M, MACMILLAN F, PRISNER T F, et al. Pulsed ENDOR at 95 GHz on the primary acceptor ubisemiquinone in photosynthetic bacterial reaction centers and related model systems[J]. J Phys Chem B, 1998, 102(23): 4648-4657.
[13] CLAUSS C, DRESSEL M, SCHEFFLER M. Optimization of coplanar waveguide resonators for ESR studies on metals[C]// J Phys Conf Ser. IOP Publishing, 2015, 592(1): 012146.
[14] PETASIS D T, HENDRICH M P. Quantitative interpretation of multifrequency multimode EPR spectra of metal containing proteins, enzymes, and biomimetic complexes[M]// Methods in Enzymology. Academic Press, 2015, 563: 171-208.
[15] WEIL J A, BOLTON J R. Electron paramagnetic resonance: elementary theory and practical applications[M]. USA: John Wiley & Sons, 2007.
[16] WEBB A. Cavity-and waveguide-resonators in electron paramagnetic resonance, nuclear magnetic resonance, and magnetic resonance imaging[J]. Prog Nucl Magn Reson Spectrosc, 2014, 83: 1-20.
[17] EARLE K A, FREED J H. Quasioptical hardware for a flexible FIR-EPR spectrometer[J]. Appl Magn Reson, 1999, 16(2): 247-272.
[18] M?BIUS K. High-field/high-frequency EPR/ENDOR—a powerful new tool in photosynthesis research[J]. Appl Magn Reson, 1995, 9: 389-407.
[19] HAINDL E, M?BIUS K. A 94 GHz EPR spectrometer with Fabry-Perot resonator[J]. Z Naturforsch A, 1985, 40(2): 169-172.
[20] EARLE K A, DZIKOVSKI B, HOFBAUER W, et al. High-frequency ESR at ACERT[J]. Magn Reson Chem, 2005, 43(S1): S256-S266.
[21] NEUGEBAUER P. Development of heterodyne high field/high frequency electron paramagnetic resonance spectrometer at 285 GHz[D]. Grenoble: Université Joseph-Fourier-Grenoble I, 2010
[22] BURGHAUS O, ROHRER M, GOTZINGER T, et al. A novel high-field/high-frequency EPR and ENDOR spectrometer operating at 3 mm wavelength[J]. Meas Sci Technol, 1992, 3(8): 765-774.
[23] BROOKER G. Modern classical optics[M]. Oxford University Press, 2003.
[24] REBUFFI L, CRENN J P. Radiation patterns of the HE11 mode and Gaussian approximations[J]. Int J Infrared Millimeter Waves, 1989, 10: 291-311.
[25] SMITH G M, LESURF J C G, MITCHELL R H, et al. Quasi-optical CW mm-wave electron spin resonance spectrometer[J]. Rev Sci Instrum, 1998, 69(11): 3924-3937.
[26] CAVALLO A, DOANE J, CUTLER R. Low-loss broadband multimode corrugated waveguide performance[J]. Rev Sci Instrum, 1990, 61(9): 2396-2400.
[27] 赵亮亮. THz金属网格带通滤波器的研究[D]. 南京: 东南大学.
[28] MATSUI T, ARAKI K, KIYOKAWA M. Gaussian-beam open resonator with highly reflective circular coupling regions[J]. IEEE Trans Microw Theory Tech, 1993, 41(10): 1710-1714.
[29] ULRICH R. Far-infrared properties of metallic mesh and its complementary structure[J]. Infrared Phys, 1967, 7(1): 37-55.
[30] HE Y, KANG R, SHI Z, et al. A design of resonant cavity with an improved coupling-adjusting mechanism for the W-band EPR spectrometer[J]. Chin Phys B, 2022, 31(11): 117601.
[31] NIELSEN R D, ROBINSON B H. The effect of field modulation on a simple resonance line shape[J]. Concept Magn Reson A, 2004, 23(1): 38-48.
[32] EATON G R, EATON S S, BARR D P, et al. Quantitative EPR[M]. Springer Science & Business Media, 2010.
[33] REED G H, MARKHAM G D. EPR of Mn (II) complexes with enzymes and other proteins[J]. Biol Magn Reson, 1984, 6: 73-142.
[34] SCHMALBEIN D, MARESCH G G, KAMLOWSKI A, et al. The Bruker high-frequency-EPR system[J]. Appl Magn Reson, 1999, 16(2): 185-205.
[35] BABUNTS R A, BADALYAN A G, GURIN A S, et al. Capabilities of compact high-frequency EPR/ESE/ODMR spectrometers based on a series of microwave bridges and a cryogen-free magneto-optical cryostat[J]. Appl Magn Reson, 2020, 51: 1125-1143.
文章导航

/