Design and Construction of a Fully-Opened Unilateral Nuclear Magnetic Resonance System

  • WANG Hong-zhi1 ,
  • 2* ,
  • LIU Xiang1 ,
  • MIAO Zhi-ying1 ,
  • CHEN Shan-shan2 ,
  • SHI Qun-yan1 ,
  • XU Luo-yuan3 ,
  • LU Zhi-yong3 ,
  • LIU Li-rong3 ,
  • YANG Pei-qiang3
Expand
  • 1. University of Shanghai for Science and Technology, Shanghai, 200093, China;
    2. Shanghai Medical Instrumentation College, Shanghai, 200093, China;
    3. Suzhou Niumag Corporation Ltd, Suzhou, 215163, China
*Corresponding author:WANG hong-zhi, Tel: 021-65485551, E-mail: wanghzhi2000@sina.com.

Received date: 2014-03-16

  Revised date: 2014-10-30

  Online published: 2014-12-05

Supported by

国家重大科学仪器设备开发专项资助项目(2013YQ170463).

摘要

为了在单边磁体产生的逸散磁场中实现核磁共振技术,信号检测深度能满足相关应用需求,该文分析了非均匀磁场下的核磁共振技术条件,确定射频场目标,设计了单边平面射频线圈和探头,搭建出一套单边核磁共振实验系统.探头具有短脉宽(90°脉宽2.5 μs),系统死时间短(10 μs)的特点,可实现15 μs 的半回波时间.应用该系统,在15.3 T/m 的巨大梯度磁场下,实现接近5 mm 厚度的信号检测,并利用单边磁体的自然梯度磁场,开展了恒定梯度磁场下的扩散测量方法探索.结果表明,该系统可满足需要一定探测深度的领域的应用需求.

本文引用格式

汪红志1 , 2* , 刘翔1 , 苗志英1 , 陈珊珊2 , 施群雁1 , 徐罗元3 , 陆治勇3 , 刘利荣3 , 杨培强3 . 全开放式单边核磁共振技术研究及系统开发[J]. 波谱学杂志, 2014 , 31(4) : 488 -503 . DOI: 10.11938/cjmr20140404

Abstract

The goal of the study is to design and construct a system that can be used to detect magnetic resonance signals generated in the stray magnetic field produced by fully-opened unilateral magnets. The radiofrequency (RF) probe was carefully designed to meet the detection depth requirement for specific applications, and tested on a unilateral nuclear magnetic resonance experiment system. The probe had a 90º pulse width as short as 2.5 s. The whole system has a short system dead time (i.e., 10 s), and the minimum half echo time was 15 s. The SGSE-CPMG experiment was performed on the system to acquire signals from a depth of 5 mm under a constant B0 gradient of 15.3 T/m. Diffusion coefficients were measured under the constant B0 gradient.

参考文献

[1] Blümich B, Haber S, Zia W. Compact NMR[M]. Berlin: De Gruyter, 2014, 1.

[2] Anferova S, Anferov V, Adams M, et al. Construction of the NMR- MOUSE with short dead time[J]. Concepts Magn Reson, 2002, 15(1): 15-25.

[3] Blümich B, Perlo J, Casanova F. Mobile single-sided NMR[J]. Nuclear Magnetic Resonance Spectroscopy, 2008, 52: 197-269.

[4] Anferova S, Anferov V, Rata D G, et al. A mobile NMR device for measurements of porosity and pore size distributions of drilled core samples[J]. Concept Magn Reson B: Magn Reson Eng, 2004, 23: 26-32.

[5] Zhang Yan-li(张艳丽), Xie De-xin(谢德馨), Bai Bao-dong(白保东), et al. Unilateral magnet for magnetic resonance imaging(单边磁体磁共振成像)[J]. Chinese J Magn Reson(波谱学杂志), 2006, 23(3): 283-292.

[6] Che Wen-hua(车文华), Xia Ping-chou(夏平畴), Zhang Yi-ming(张一鸣). Discussion on design of anti-traditional MRI magnet system(关于反传统核磁共振成像装置磁体系统设计的探讨)[J]. Electrotechnical Journal(电工技术杂志), 2000, 9: 23-25.

[7] Xie Jun-peng(谢俊鹏), Yao Ying-ying(姚缨英). Design of completely open MRI magnet(完全开放方式磁共振成像磁体的设计)[D]. Zhejiang University master's degree thesis(浙江大学硕士学位论文), 2006.

[8] Liu H B, Xiao L Z, Guo B X, et al. Heavy oil component characterization with multidimensional unilateral NMR[J]. Petroleum Science, 2013, 10: 402-407.

[9] Xu Zheng(徐征), Guo Pan(郭盼), He Xiao-long(何晓龙), et al. Based on the principles of NMR study of the aging of composite insulators(基于核磁共振原理的复合绝缘子老化问题研究)[J]. High-Voltage Electrical(高压电器), 2012, 48(3): 21-25.

[10] Watzlaw J, Müntjes J, Mokwa W, et al. Mobile single-sided NMR with micro structured multilayer coil arrays[C]. 53rd Experimental Nuclear Magnetic Resonance Conference, Poster 174.

[11] Chang W H, Chung C Y, Chen J H, et al. Simple mobile single-sided NMR apparatus with a relatively homogeneous B0 distribution[J]. MRI, 2011, 29: 869-876.

[12] Maxime L, Ernesto D, Juan P, Blümich B. Single-sided magnetic resonance profiling in biological and materials science[J]. J Magn Reson, 2012, 215: 74-84.

[13] Wei Shu-feng(魏树峰), Yang Wen-hui(杨文晖), Guo Li-hui(郭丽慧), et al. RF transmitter coil design for open MRI(一种用于开放式MRI的射频发射线圈设计)[J]. Chinese J Magn Reson(波谱学杂志), 2012, 29(3): 354-360.

[14] Blümich B, Casanova F, Appelt S, et al. NMR at low magnetic fields[J]. Chem Phys Lett, 2009, 477: 231-240.

[15] Ge Cheng(葛成), Guo Xin-jun(郭新军), Zhao Hai-bo(赵海波), et al. Factors of MRI T2 inversion(核磁共振T2谱反演影响因素)[J]. Geophysical and Geochemical Exploration(物探与化探), 2011, 35(6):845-854.

[16] Juan C, Igor M, Bruce C, et al. A unilateral magnet with an extended constant magnetic field gradient[J]. J Magn Reson, 2010, 207: 337-344.

[17] Pulyer Y M. Planar open sole-noidal magnet MRI system US Patent, 5914600 [P], 1999-06-22.

[18] Li Peng-ju(李鹏举). The study of Fluid recognition method and NMR T2 spectral inversion(核磁共振T2谱反演及流体识别方法研究)[D]. Northeast Petroleum University(东北石油大学), 2010.

文章导航

/