Design and Implementation of An Active Shimming System for Low Field NMR Analyzers

  • CHEN Shan-shan ,
  • YANG Pei-qiang ,
  • WANG Hong-zhi ,
  • BI Xin ,
  • XIA Tian ,
  • MIAO Zhi-ying ,
  • DAI Shu-guang
Expand
  • 1. Scool of Optical-Electrical and Computer Engineering, University of Shanghai for Science and Technology, Shanghai, 200093, China;
    2. Shanghai University of Medicine & Health Science, Shanghai, 201318, China;
    3. Suzhou Niumag Corporation Ltd, Suzhou 215163, China

Received date: 2015-12-21

  Revised date: 2017-01-04

  Online published: 2017-03-05

Abstract

High homogeneity of the magnet field is an essential prerequisite for measuring weak signals and short spin-spin relaxation time on a nuclear magnetic resonance (NMR) analyzer. In this study, we designed active shimming coils, based on a target-field approach, for low filed NMR analyzers. Constant current source was designed according to the coil power supply. An active shimming platform used in a NMR analyzer's 0.45 T magnet system was constructed over a large target-diameter-sphere-volume (25.4 mm DSV) of interest. The basic structure and operation principles of the system, as well as the shimming method, were introduced. The feasibility and practicality of the design was demonstrated by experimental measurements.

Cite this article

CHEN Shan-shan , YANG Pei-qiang , WANG Hong-zhi , BI Xin , XIA Tian , MIAO Zhi-ying , DAI Shu-guang . Design and Implementation of An Active Shimming System for Low Field NMR Analyzers[J]. Chinese Journal of Magnetic Resonance, 2017 , 34(1) : 87 -99 . DOI: 10.11938/cjmr20170111

References

[1] 裘祖文, 裴奉奎. 核磁共振波谱[M]. 北京:科学出版社, 1989.
[2] SONG Y Q. Magnetic resonance of porous media (MRPM):A perspective[J]. J Magn Reson, 2013, 229:12-24.
[3] SONG Y Q. Low field magnetic resonance:Multi-dimensional experiments of relaxation and diffusion[J]. Chinese J Magn Reson, 2015, 32(2):141-149. 宋一桥. 低场核磁共振:弛豫和扩散的多维实验[J]. 波谱学杂志, 2015, 32(2):141-149.
[4] ZHOU S W, XUE H Q, GUO W, et al. Measuring movable oil saturation in reservoirs with low-field NMR technology[J]. Chinese J Magn Reson, 2015, 32(3):489-498. 周尚文, 薛华庆, 郭伟, 等. 基于低场核磁共振技术的储层可动油饱和度测试新方法[J]. 波谱学杂志, 2015, 32(3):489-498.
[5] CHEN L M, PENG C L, ZHU X W. The development of an active shimming system for 0.2T open-style permanent magnet MRI[J]. J Biomed Eng, 2004, 21(2):288-291. 陈历明, 彭承琳, 朱学武. 0.2T开放式永磁型医用核磁共振成像仪有源匀场系统的研制[J]. 生物医学工程学杂志, 2004, 21(2):288-291.
[6] CHU Y S. Active shimming technique of NMR imaging magnet[J]. Transactions of China Electrotechnical Society, 1996, 11(5):40-44. 储岳森. 核磁共振成像磁体的有源匀场技术[J]. 电工技术学报, 1996, 11(5):40-44.
[7] WU H C, LIU Z M, ZHOU H Q. The design of lower-order tabulate shim coils for MRI system[J]. Proceedings of the CSEE, 2005, 25(24):154-159. 武海澄, 刘正敏, 周荷琴. 磁共振成像系统中低阶平板式匀场线圈的设计[J]. 中国电机工程学报, 2005, 25(24):154-159.
[8] WANG H Z, JIANG X, ZHANG X L, et al. Design of active shimming coils on mini-type permanent magnetic resonance imaging system[J]. J Biomed Eng, 2009, 26(3):465-470. 汪红志, 蒋欣, 张学龙, 等. 永磁微型核磁共振成像系统匀场线圈设计[J]. 生物医学工程学杂志, 2009, 26(3):465-470.
[9] ZHENG Y H, GAO H B. Magnetic field homogeneity in high resolution superconducting NMR spectrometer[J]. Chinese J Magn Reson, 1985, 2(4):393-411. 郑耀华, 高汉宾. 高分辨超导NMR谱仪的匀场. 波谱学杂志, 1985, 2(4):393-411.
[10] WANG Q Y. Correcting magnetic field inhomogeneities of a permanent magnet applied to NMR imaging[J]. Advanced Technology of Electrical Engineering and Energy, 1989, 8(4):42-49. 王清源. 核磁共振成像用永磁体不均匀场的调整[J]. 电工电能新技术, 1989, 8(4):42-49.
[11] LAWRENCE K F, STUART C. A novel target-field method for magnetic resonance shim coils:Ⅲ. Shielded zonal and tesseral coils[J]. Journal of Physics D:Applied Physics, 2003, 36(2):68-80.
[12] LIU W T, TANG X, ZU D L. A novel target-field approach to design bi-planar shim coils for permanent-magnet MRI[J]. Concep Magn Reson B, 2010, 37B(1):29-38.
[13] FORBES L K, BRIDESON M A, CROZIER S. A target-field method to design circular biplanar coils for asymmetric shim and gradient fields[J]. IEEE T Magn, 2005, 41(6):2134-2144.
[14] 王玉. 核磁共振室温匀场电源与场频联锁系统一体化——核心板高速电路设计[D]. 厦门:厦门大学, 2012
[15] 谈军. 核磁共振波谱仪匀场电源数据通信模块设计[D]. 厦门:厦门大学, 2009
[16] LI J, CHEN H B, TAN J, et al. An FPGA implementation of the control system for shim power in room temperature[J]. Journal of Xiamen University (Natural Science), 2010, 49(3):319-323. 李洁, 陈海波, 谈军, 等. 基于FPGA的NMR谱仪室温匀场电源核心控制系统的设计[J]. 厦门大学学报(自然科学版), 2010, 49(3):319-323.
[17] 鲍庆嘉. 高分辨核磁共振实验与数据处理的自动化关键技术研究[D]. 武汉:华中科技大学, 2013.
[18] WANG H W, LI Z G, MAO W P, et al. The design of integrated gradient -field/frequency lock-shim system for NMR spectrometer[J]. Chinese J Magn Reson, 2012, 29(4):489-498. 王恢旺, 李正刚, 毛文平, 等. 一体化NMR波谱仪梯度-场频联锁-匀场系统设计[J]. 波谱学杂志, 2012, 29(4):489-498.
[19] LI J S, CHEN Y S. The design and manufacturing of golay shimming coils for NMR[J]. Analytical Instrumentation, 1983, 14(4):27-37. 李杰森, 陈应书. 用于核磁共振仪的高雷型匀场线圈的设计和制造[J]. 分析仪器, 1983, 14(4):27-37.
[20] ZHAO W, LIU Z W, LIU Y, et al. Shimming method for permanent of open-type-field MRI[J]. Advanced Technology of Electrical Engineering and Energy, 2007, 26(3):29-32. 赵微, 刘志文, 刘勇, 等. 开放式低场永磁MRI磁体的匀场方法研究[J]. 电工电能新技术, 2007, 26(3):29-32.
[21] TANG X Y, LIU Z W, ZHAO W. Gradient shimming in magnetic resonance imaging magnet[J]. Transactions of Beijing Institute of Technology, 2009, 29(3):279-282.
Outlines

/