High Accuracy Generation and Test of B0 Signals on Magnetic Resonance Imaging Spectrometer

  • CUI Yan ,
  • XIAO Liang
Expand
  • College of Information Science and Technology, Beijing University of Chemical Technology, Beijing 100029, China

Received date: 2017-09-06

  Online published: 2017-11-28

Abstract

A field programmable gate array (FPGA)-based method is proposed to generate B0 signals, with high accuracy, on magnetic resonance imaging (MRI) spectrometers. The FPGA reads the waveform data and external parameters, and stores them in the dual-port random access memory (RAM) and parameter registers, respectively. The B0 signals are then generated using the pre-emphasis algorithm, and different time or amplitude parameters give different B0 signals. The B0 signals are collected by high speed acquisition card, and the least squares fitting algorithm is used to test the accuracy of the B0 signals generated. The performance of this method was validated by experiments.

Cite this article

CUI Yan , XIAO Liang . High Accuracy Generation and Test of B0 Signals on Magnetic Resonance Imaging Spectrometer[J]. Chinese Journal of Magnetic Resonance, 2018 , 35(2) : 170 -177 . DOI: 10.11938/cjmr20172594

References

[1] LIU M, QIU W Q, Sun H J, et al. Progress in the portable NMR spectrometer[J]. Chinese J Magn Reson. 2014, 31(4):504-514. 刘敏, 邱雯琦, 孙慧军, 等. 便携式核磁共振谱仪的研究进展[J]. 波谱学杂志, 2014, 31(4):504-514.
[2] SHEN J, LIU Y, LI J Q, et al. A powerful graphical pulse sequence programming tool for magnetic resonance imaging[J]. MAGMA, 2005, 18(6):332-342.
[3] MAO W P, BAO Q J, YANG L, et al. A modularized pulse programmer for NMR spectroscopy[J]. Meas Sci Technol, 2011, 22(2):025901.
[4] STANG P P, CONOLLY S M, SANTOS J M, et al. Medusa:a scalable MR console using USB[J]. IEEE Trans Med Imaging, 2012, 31(2):370-379.
[5] HOU S L, XIE H T, HOU X W, et al. Gradient coil of permanent magnet mini-magnetic resonance imager and image quality[J]. Chinese J Magn Reson. 2014, 29(4):508-520. 侯淑莲, 谢寰彤, 侯晓吻, 等. 永磁微型磁共振成像仪的梯度线圈与图像质量[J]. 波谱学杂志, 2014, 29(4):508-520.
[6] XU Q, WANG H, JIANG Y, et al. A digital preemphasis gradient waveform generator for magnetic resonance imaging[J]. Chinese J Magn Reson, 2006, 23(1):11-16. 徐勤, 王鹤, 蒋瑜, 等. 一种具有数字预加重的磁共振成像梯度波形发生器[J]. 波谱学杂志, 2006, 23(1):11-16.
[7] XIAO L, TANG W N, WANG W M. An FPGA-based single-chip gradient calculation module for magnetic resonance imaging[J]. Chinese J Magn Reson, 2010, 27(2):163-171. 肖亮, 汤伟男, 王为民. 基于单片FPGA的磁共振成像梯度计算模块[J]. 波谱学杂志, 2010, 27(2):163-171.
[8] ZANG F C, LI J Q, WANG H, et al. Automatic adjustment of gradient preemphasis in MRI systems[J]. Chinese J Magn Reson, 2008, 25(1):26-32. 臧凤超, 李建奇, 王鹤, 等. 磁共振成像系统中的自动梯度预加重调节方法研究[J]. 波谱学杂志, 2008, 25(1):26-32.
[9] LIU H D, WANG H W, ZHU T X, et al. A frequency synthesizer for high-field NMR spectrometers[J]. Chinese J Magn Reson, 2014, 31(1):91-99. 刘海丹, 王辉旺, 朱天雄, 等. 高场核磁共振波谱仪频率合成器设计[J]. 波谱学杂志, 2014, 31(1):91-99.
[10] FRY M E, PITTARD S, SUMMERS I R, et al. A programmable eddy-current compensation system for MRI and localized spectroscopy[J]. J Magn Reson Imaging, 1997, 7(2):455-458.
[11] VARDAL J, SALO R A, LAESSON C, et al. Correction of B0-distortions in echo-planar-imaging-based perfusion-weighted MR[J]. J Magn Reson Imaging, 2014, 39(3):722-728.
Outlines

/