Gradient coils generate gradient magnetic fields that are essential in the magnetic response imaging. The gradient coils are usually designed by the target field method. In this work, we attempted to design matrix gradient coils with particle swarm optimization and the genetic algorithm. Finite element simulation calculation demonstrated that the average non-uniformity in the x and y directions was 0.851%, and the average non-uniformity in the z direction was 1.013%.
WANG Qiang
,
WEI Shu-feng
,
WANG Zheng
,
YANG Wen-hui
. Design of Matrix Gradient Coils with Particle Swarm Optimization and the Genetic Algorithm[J]. Chinese Journal of Magnetic Resonance, 2019
, 36(4)
: 463
-471
.
DOI: 10.11938/cjmr20192730
[1] TURNER R. Gradient coil design:A review of methods[J]. Magn Reson Imaging, 1993, 11(7):903-920.
[2] TURNER R. A target field approach to optimal coil design[J]. J Phys D Appl Phys, 1986, 19(8):L147-L151.
[3] MANSFIELD P, CHAPMAN B L, BOWTELL R, et al. Active acoustic screening:reduction of noise in gradient coils by Lorentz force balancing[J]. Magn Reson Med, 1995, 33(2):276-281.
[4] YODA K. Analytical design method of self-shielded planar coils[J]. J Appl Phys, 1990, 67(9):4349-4353.
[5] CHRISTOPH J, NIXON T W, SCOTT M I, et al. Magnetic field modeling with a set of individual localized coils[J]. J Magn Reson, 2010, 204(2):281-289.
[6] JIA F, SCHULTZ G, TESTUD F, et al. Performance evaluation of matrix gradient coils[J]. MAGMA, 2016, 29(1):59-73.
[7] JIA F, LITTIN S, LAYTON K J, et al. Design of a shielded coil element of a matrix gradient coil[J]. J Magn Reson, 2017, 281:217-228.
[8] KROBOTH S, LAYTON K J, JIA F, et al. Optimization of coil element configurations for a matrix gradient coil[J]. IEEE T Med Imaging, 2018, 37(1):284-292
[9] SHI X H, WAN L M, LEE H P, et al. An improved genetic algorithm with variable population-size and a PSO-GA based hybrid evolutionary algorithm[C]//International Conference on Machine Learning & Cybernetics, 2003. doi:10.1109/ICMLC.2003.1259777.