Review Articles

Matrix Gradient Coil: Current Research Status and Perspectives

  • HE Hong-yan ,
  • WEI Shu-feng ,
  • WANG Hui-xian ,
  • YANG Wen-hui
Expand
  • 1. Institute of Electrical of Engineering, Chinese Academy of Sciences, Beijing 100190, China;
    2. University of Chinese Academy of Sciences, Beijing 100049, China

Received date: 2020-03-05

  Online published: 2020-05-22

Abstract

The performance of gradient and shim coils affect the quality of magnetic resonance imaging (MRI) directly. Conventional gradient/shim coils have some inherent disadvantages, such as the state of produced magnetic field is single and inflexible, many types of coils are required and the structure is complex, while the matrix gradient coils are able to overcome those problems. This paper introduces the concept and characteristics of matrix gradient coil, and reviews the current progresses in this research topic according to the structure and function of the gradient coil. The development tendency of matrix gradient coil in the future is proposed. In addition, the preliminary research on matrix gradient coil is presented.

Cite this article

HE Hong-yan , WEI Shu-feng , WANG Hui-xian , YANG Wen-hui . Matrix Gradient Coil: Current Research Status and Perspectives[J]. Chinese Journal of Magnetic Resonance, 2021 , 38(1) : 140 -153 . DOI: 10.11938/cjmr20202812

References

[1] GOLAY, MARCEL J E. Field homogenizing coils for nuclear spin resonance instrumentation[J]. Rev Sci Instrum, 1958, 29(4):313-315.
[2] SIEBOLD H. Gradient field coils for MR imaging with high spectral purity[J]. IEEE T Magn, 1990, 26(2):897-900.
[3] TURNER R. A target field approach to optimal coil design[J]. J Phys D Appl Phys, 1986, 19(8):L147-L151.
[4] TURNER R. Minimum inductance coils[J]. J Phys E Sci Instrum, 1988, 21(10):948-952.
[5] PEEREN G N. Stream function approach for determining optimal surface currents[J]. J Comput Phys, 2003, 191(1):305-321.
[6] JUCHEM C, BIERL B M, SCHICK F, et al. Combined passive and active shimming for in vivo MR spectroscopy at high magnetic fields[J]. J Magn Reson, 2006, 183:278-289.
[7] JUCHEM C, BROWN P B, NIXON T W, et al. Multicoil shimming of the mouse brain[J]. Magn Reson Med, 2011, 66(3):893-900.
[8] ZHOU J Z, STOCKMANN J P, ARANGO N, et al. The orthogonal shim coil for 3T brain imaging[J]. Magn Reson Med, 2019, 83(1):1-13.
[9] WINKLER S A, SCHMITT F, LANDES H, et al. Gradient and shim technologies for ultra high field MRI[J]. Neuroimage, 2018, 168:59-70.
[10] HARRIS C T, HANDLER W B, CHRONIK B A. A new approach to shimming:The dynamically controlled adaptive current network[J]. Magn Reson Med, 2014, 71(2):859-869.
[11] JUCHEM C, NIXON T W, MCINTYRE S, et al. Magnetic field modeling with a set of individual localized coils[J]. J Magn Reson, 2010, 204(2):281-289.
[12] JUCHEM C, NIXON T W, MCINTYRE S, et al. Dynamic multi-coil shimming of the human brain at 7 T[J]. J Magn Reson, 2011, 212(2):280-288.
[13] JUCHEM C, NIXON T W, GRAAF R A D. Multi-coil imaging with algebraic reconstruction[C]//Proceedings of the 20th Annual Meeting of ISMRM, Melbourne, Australia, 2012. Abstract:2545.
[14] JUCHEM C, GRAAF R A D. The public multi-coil information (PUMCIN) policy[J]. Magn Reson Med, 2017, 78(5):2042-2047. doi:10.1002/mrm.26558.
[15] WINTZHEIMER S, DRIESSLE T, LEDWIG M, et al. A 50-channel matrix gradient system:a feasibility study[C]. In Proceedings of the joint annual meeting ISMRM-ESMRMB, Stockholm, Sweden, 2010:3937
[16] JIA F, LITTIN S, LAYTON K J, et al. Design of a shielded coil element of a matrix gradient coil[C]//Proc Intl Soc Mag Reson Med, 23, 2015:3091
[17] WHILE P T, KORVINK J G. Designing MR shim arrays with irregular coil geometry:theoretical considerations[J]. IEEE T Biomed Eng, 2014, 61(6):1614-1620.
[18] JUCHEM C, NAHHASS O M, NIXON T W, et al. Multi-slice MRI with the dynamic multi-coil technique[J]. NMR Biomed, 2015, 28(11):1526-1534.
[19] YANG S G, LU X F, FU J Q, et al. Safety alert of high field or ultra-high field MRI for using[J]. Biomedical Engineering and Clinical Medicine, 2014, 18(1):94-98.杨述根,陆笑非,付家庆,等.高场或超高场MRI使用安全警示[J].生物医学工程与临床, 2014, 18(1):94-98.
[20] TIAN J G, LIU M L, XIA Z F, et al. Safety of magnetic resonance imaging[J]. Chinese J Magn Reson, 2000, 17(6):505-511.田建广,刘买利,夏照帆,等.磁共振成像的安全性[J].波谱学杂志, 2000, 17(6):505-511.
[21] GLEICH B, WEIZENECKER J. Tomographic imaging using the nonlinear response of magnetic particles[J]. Nature, 2005, 435(7046):1214-1217.
[22] JIA F, LITTIN S, LAYTON K J, et al. Design of a shielded coil element of a matrix gradient coil[J]. J Magn Reson, 2017, 281:217-228.
[23] TOPFER R, STAREWICZ P, LO K M, et al. A 24-channel shim array for the human spinal cord:design, evaluation, and application[J]. Magn Reson Med, 2016, 76(5):1604-1611.
[24] SONG R, HE Y F, ZHANG B. A method for eddy current field measurement in permanent magnet magnetic resonance imaging systems[J]. Chinese J Magn Reson, 2018, 35(1):52-59.宋瑞,何砚发,张波.一种永磁磁共振中磁体涡流场的测量方法[J].波谱学杂志, 2018, 35(1):52-59.
[25] WONG E C, MAZAHERI Y. Shimming of the inferior frontal cortex using an external local shim coil[J]. Proc Intl Soc Mag Reson Med, 2004, 11:520.
[26] HSU J J, GLOVER G H. Mitigation of susceptibility-induced signal loss in neuroimaging using localized shim coils[J]. Magn Reson Med, 2005, 53(2):243-248.
[27] JUCHEM C, NIXON T W, MCINTYRE S, et al. Magnetic field homogenization of the human prefrontal cortex with a set of localized electrical coils[J]. Magn Reson Med, 2010, 63(1):171-180.
[28] LIU J C, HUANG K M, HU Y Y, et al. Design and field calculation of coil array for functional magnetic stimulation[J]. Space Medicine & Medical Engineering, 2004, 5:365-369.刘冀成,黄卡玛,胡雅毅,等.功能磁刺激线圈阵列设计与场分布计算[J].航天医学与医学工程, 2004, 5:365-369.
[29] HENNIG J, WELZ A M, SCHULTZ G, et al. Parallel imaging in non-bijective, curvilinear magnetic field gradients:a concept study[J]. MAGMA, 2008, 21(1,2):5-14.
[30] LIN F H, WITZEL T, SCHULTZ G, et al. Reconstruction of MRI data encoded by multiple nonbijective curvilinear magnetic fields[J]. Magn Reson Med, 2012, 68(4):1145-1156.
[31] JUCHEM C, HERMAN P, SANGANAHALLI B G, et al. DYNAmic Multi-coIl TEchnique (DYNAMITE) shimming of the rat brain at 11.7Tesla[J]. NMR Biomed, 2014, 27(8):897-906.
[32] JUCHEM C, NIXON T W, BROWN P B, et al. Spatial selection through multi-coil magnetic field shaping[C]//Proceedings of the 19th Annual Meeting of ISMRM, Montreal, Canada:2011. Abstract 385.
[33] JUCHEM C, GREEN D, GRAAF R A D. Multi-coil magnetic field modeling[J]. J Magn Reson, 2013, 236C:95-104.
[34] JIA F, SCHULTZ G, TESTUD F, et al. Performance evaluation of matrix gradient coils[J]. MAGMA 29, 2016:59-73.
[35] AGHAEIFAR A, MIRKES C, BAUSE J, et al. Dynamic B0 shimming of the human brain at 9.4 T with a 16-channel multi-coil shim setup[J]. Magn Reson Med, 2018:1-12.
[36] RUDRAPATNA S, FLUERENBROCK F, NIXON T W, et al. Combined imaging and shimming with the dynamic multi-coil technique[J]. Magn Reson Med, 2019, 81(2):1424-1433.
[37] 中国科学院电工研究所.杨文晖,王铮.一种磁共振成像系统的梯度磁场产生方法及其装置:中国. 105301536 A[P], 2016-02-03.
[38] 中国科学院电工研究所.杨文晖,魏树峰,王铮,等.磁共振成像系统的阵列梯度线圈:中国. 107219480 A[P], 2017-09-29.
[39] WANG Q, WEI S F, WANG Z, et al. Design of matrix gradient coils with particle swarm optimization and the genetic algorithm[J]. Chinese J Magn Reson, 2019, 36(4):463-471.王强,魏树峰,王铮,等.基于粒子群与遗传算法的矩阵式梯度线圈优化设计[J].波谱学杂志, 2019, 36(4):463-471.
[40] 徐雅洁,郁朋,吴振洲,等. Halbach阵列磁体中阵列线圈主动匀场技术[C]//2018第二十届全国波谱学学术年会会议论文摘要集.中国物理学会波谱专业委员会, 2018:326-327.
[41] LITTIN S, JIA F, LAYTON K J, et al. Development and implementation of an 84-channel matrix gradient coil[J]. Magn Reson Med, 2018, 79(2):1181-1191.
[42] STOCKMANN J P, WALD L L. In vivo B0 Field shimming methods for MRI at 7 T[J]. NeuroImage, 2017, 168:71-87.
[43] HAN H, SONG A W, TRUONG T K. Integrated parallel reception, excitation, and shimming (iPRES)[J]. Magn Reson Med, 2013, 70(1):241-247.
[44] STOCKMANN J P, WITZEL T, BLAU J, et al. Combined shim-RF array for highly efficient shimming of the brain at 7 Tesla[C]//Proceedings of the 21st Annual Meeting of ISMRM; Salt Lake City, Utah, USA:2013:665.
[45] GAO Y, MAREYAM A, SUN Y, et al. A 16-channel AC/DC array coil for anesthetized monkey whole-brain imaging at 7 T[J]. NeuroImage, 2020, 207:116396.
[46] WINKLER S A, WARR P A, STOCKMANN J P, et al. Comparison of new element designs for combined RF-shim arrays at 7 T[J]. Concepts Magn Reson B, 2018:e21364.
[47] DARNELL D, TRUONG T K, SONG A W, et al. Integrated parallel reception, excitation, and shimming (iPRES) with multiple shim loops per RF coil element for improved B0 shimming[J]. Magn Reson Med, 2017, 77(5):2077-2086.
[48] STOCKMANN J P, WITZEL T, KEIL B, et al. A 32-channel combined RF and B0 shim array for 3 T brain imaging[J]. Magn Reson Med, 2016, 75(1):441-451.
[49] KROBOTH S, LAYTON K J, JIA F, et al. Optimization of coil element configurations for a matrix gradient coil[J]. IEEE T Med Imaging, 2018, 37(1):284-292.
[50] STOCKMANN J P, WITZEL T, KEIL B, et al. A 32ch combined RF-shim brain array for efficient B0 shimming and RF reception at 3 T[C]//Proceedings of the 22nd Annual Meeting of ISMRM, Milan. Italy:2014:400.
[51] DARELL D, MA Y X, WANG H Y, et al. Adaptive integrated parallel reception, excitation, and shimming (iPRES-A) with microelectromechanical systems switches[J]. Magn Reson Med, 2018, 80(1):371-379
[52] AGHAEIFAR A, ZHOU J Z, HEULE R, et al. A 32-channel multi-coil setup optimized for human brain shimming at 9.4 T[J]. Magn Reson Med, 2020, 83(2):749-764.
[53] CHEN W J, CHEN J, SUN H J, et al. Simulation and analysis of irregular multicoil B0 shimming in C-type permanent magnets using genetic algorithm and simulated annealing[J]. Appl Magn Reson, 2019, 50(1-3):227-242.
[54] ERTAN K, TARAGHINIA S, ATALAR E. Driving mutually coupled gradient array coils in magnetic resonance imaging[J]. Magn Reson Med, 2019, 82(3):1187-1198.
[55] "十三五"医疗器械科技创新专项规划.科技部国科办社[2017] 44号.[2017.5.26]. http://www.most.gov.cn/tztg/201706/t20170612_133477.htm.
Outlines

/