Chinese Journal of Magnetic Resonance >
Optimization Analysis and Experimental Verification of FID NMR Coil in Polarized 3He Systems
Received date: 2024-02-26
Online published: 2024-03-26
The application of hyperpolarized 3He gas is extensive in fields such as neutron polarimetry, fundamental physics, and diagnostic medicine. The polarimetry measurement is crucial for 3He polarization systems. Typically, a free induction decay nuclear magnetic resonance (FID NMR) coil is used to detect the FID signal of 3He atoms. The measurement signal-to-noise ratio (SNR) of the coil is correlated with factors such as diameter, number of turns, and relative position to the cell. There is limited literature and insufficient experimental verification on the optimization analysis of FID NMR coils. In this study, an SNR model for the coil was established through theoretical analysis of the FID signal and coil noise. The FID NMR coil was experimentally validated in a 3He polarization system based on the metastability-exchange optical pumping (MEOP) technique, which is known for its high polarization efficiency. The experimental results indicated that for a cylindrical cell with both radius and height of rcell, the optimal SNR of the coil was achieved at a radius of rcell/2, which was consistent with the theoretical analysis. The relative error of the fitted SNR model was less than 10%, demonstrating its effectiveness. These research findings hold significant practical implications for optimizing the design of 3He polarization systems.
Key words: FID coil design; polarimetry measurement; polarized 3He; MEOP
LUO Qingjin , WU Liangyong , WANG Yuting , YAN Haiyang , XIANG Yifeng , CHEN Siyu . Optimization Analysis and Experimental Verification of FID NMR Coil in Polarized 3He Systems[J]. Chinese Journal of Magnetic Resonance, 2024 , 41(3) : 266 -275 . DOI: 10.11938/cjmr20243099
| [1] | WALKER T G. Fundamentals of spin-exchange optical pumping[J]. J Phys: Conf Ser, 2011, 294: 012001. |
| [2] | YAN S, ZHANG M F, GUO W C, et al. Development of a polarized 3He neutron spin filter based on spin exchange optical pumping at China Mianyang research reactor[J]. Sci China Phys Mech, 2019, 62(10): 1-4. |
| [3] | GENTILE T R, NACHER P J, SAAM B, et al. Optically polarized 3He[J]. Rev Mod Phys, 2017, 89(4): 045004. |
| [4] | JIANG C Y. An overview of polarized neutron instruments and techniques in Asia pacific[J]. AAPPS Bulletin, 2023, 33: 21. |
| [5] | WILD J M, STEWART N J, CHAN H F. Hyperpolarised helium-3 (3He) MRI: physical methods for imaging human lung function[M]// KAUCZOR H, WIELPüTZ M O. MRI of the lung. Medical Radiology. Springer, Cham, 2017: 69-97. |
| [6] | LEE W T, ZHENG G, TALBOT C L. Hyperpolarised gas filling station for medical imaging using polarised 129Xe and 3He[J]. Magn Reson Imaging. 2021, 79: 112-120. |
| [7] | CHEN S Z, LAN Y, LI H D, et al. Relationship between lung and brain injury in covid-19 patients: A hyperpolarized 129Xe-MRI-based 8-month follow-up[J]. Biomedicines, 2022, 10(4): 781-790. |
| [8] | LI H D, ZHAO X C, WANG Y J, et al. Damaged lung gas exchange function of discharged covid-19 patients detected by hyperpolarized 129Xe MRI[J]. Sci Adv, 2021, 7(1): eabc8180. |
| [9] | FAROOQ M, CHUPP T, GRANGE J, et al. Absolute magnetometry with 3He[J]. Phys Rev Lett, 2020, 124: 223001. |
| [10] | NIKIEL A, BL’UMLER P, HEIL W, et al. Ultrasensitive 3He magnetometer for measurements of high magnetic fields[J]. Eur Phys J D, 2014, 68(11): 330. |
| [11] | YAN H, SUN G A, PENG S M, et al. Searching for new spin- and velocity-dependent interactions by spin relaxation of polarized 3He gas[J]. Phys Rev Lett, 2015,115: 182001. |
| [12] | YAN H, SUN G A, GONG J, et al. Probing the short range spin dependent interactions by polarized 3He atom beams[J]. Eur Phys J C, 2014,74: 3088. |
| [13] | CHU P H, DENNIS A, FU C B, et al. Laboratory search for spin-dependent short-range force from axionlike particles using optically polarized 3He gas[J]. Phys Rev D, 2013, 87: 011105. |
| [14] | ZHENG W, GAO H, LALREMRUATA B, et al. Search for spin-dependent short-range force between nucleons using optically polarized 3He gas[J]. Phys Rev D, 2012, 85: 031505. |
| [15] | PETUKHOV A K, PIGNOL G, JULLIEN D, et al. Polarized 3He as a probe for short-range spin-dependent interactions[J]. Phys Rev Lett, 2010, 105: 170401. |
| [16] | WANG C F, LAN Y, YAN H Y, et al. Laser detection of nuclear magnetic resonance FID signal for polarized 3He system[J]. Chinese J Magn Reson, 2022, 39(4): 459-466. |
| 王昶沣, 兰阳, 闫海洋, 等. 极化3He系统核磁共振FID信号的激光探测[J]. 波谱学杂志, 2022, 39(4): 459-466. | |
| [17] | TU X Q, ZHENG H, SUN G A, et al. Pickup coil optimization for polarized 3He system[J]. Eur Phys J Plus, 2017, 132: 308. |
| [18] | 闫松. 3He极化率测量系统设计、实现与优化[D]. 北京: 北京交通大学, 2019. |
| [19] | 屠小青. 极化3He中子极化系统研制及其在中子散射中的应用研究[D]. 绵阳: 中国工程物理研究院, 2017. |
| [20] | YAN S, TU X Q, PENG M. Design and implementation of a monitoring system for measuring polarization degree of optical pumping 3He[J]. Chinese J Magn Reson, 2020, 37(1): 114-122. |
| 闫松, 屠小青, 彭梅. 光泵抽运3He极化程度监控系统的设计与实现[J]. 波谱学杂志, 2020, 37(1): 114-122. | |
| [21] | CACIAGLI A, BAARS R J, PHILIPSE A P, et al. Exact expression for the magnetic field of a finite cylinder with arbitrary uniform magnetization[J]. J Magn Magn Mater, 2018, 456: 423-432. |
| [22] | 高晋占. 微弱信号检测[M]. 第三版. 北京: 清华大学出版社, 2019. |
/
| 〈 |
|
〉 |