基于流形结构正则化的快速高质量磁共振指纹定量成像
李鹏, 纪雨萍, 胡悦

High-quality MR Fingerprinting Reconstruction Based on Manifold Structured Data Priors
LI Peng, JI Yuping, HU Yue
表2 不同方法在使用不同采集长度L时重建参数图的归一化均方误差(NMSE)
Table 2 The NMSEs of the reconstructed parameter maps by different methods using various acquisition lengths L
L 200 300 400 500
Maps T1 T2 PD T1 T2 PD T1 T2 PD T1 T2 PD
BLIP 0.0872 0.3789 0.0321 0.0682 0.2830 0.0311 0.0564 0.2080 0.0302 0.0453 0.1549 0.0291
MBIR 0.0345 0.3490 0.0369 0.0344 0.2310 0.0353 0.0328 0.1524 0.0347 0.0311 0.1160 0.0318
FLOR 0.0208 0.1676 0.0190 0.0143 0.1067 0.0131 0.0129 0.0680 0.0128 0.0104 0.0450 0.0101
SL-SP 0.0163 0.1417 0.0108 0.0112 0.0967 0.0105 0.0108 0.0667 0.0081 0.0102 0.0429 0.0075
DG-LR 0.0167 0.1421 0.0106 0.0109 0.1013 0.0104 0.0103 0.0659 0.0073 0.0075 0.0365 0.0045
MS 0.0219 0.1901 0.0180 0.0149 0.0991 0.0097 0.0137 0.0653 0.0082 0.0093 0.0406 0.0048
LLR 0.0186 0.1890 0.0132 0.0134 0.1117 0.0091 0.0120 0.0659 0.0088 0.0081 0.0400 0.0046
MS-LLR(本文方法) 0.0146 0.1374 0.0080 0.0094 0.0864 0.0046 0.0078 0.0603 0.0040 0.0052 0.0284 0.0027