Abstract: On the hypothesis of infinite length cylindrical medium,demagnetization factor, geometrical factor and susceptibility correction formula for arbitrary angular orientation of NMR sample tube with reference to the magnetic field are derived. It is concluded that under magic angle condition, solution susceptibility effect could be eliminated, which is similar to that of ellipsoid medium sample by Garroway. Exprimental results agree well with the theoretical deduction obtained.
Mao Xian, Shen Lianfang, Wu Qinyi. DEMAGNETIZATION FACTOR, GEOMETRICAL FACTOR AND SUSCEPTIBILITY CORRECTION FORMULA FOR CYLINDRICAL SAMPLE UNDER UNIVERSAL CONDITIONS[J]. Chinese Journal of Magnetic Resonance, 1985, 2(3): 249-252.