DEVELOPMENT OF L-BAND THREE DIMENSIONAL ESR IMAGING SYSTEM(Ⅱ)——L-Band Three Dimensional ESR Imaging System

ZHENG Ying-guang1,  XU JING2, DONG Feng-xia1,  SHEN Er-zhong3, XU Yu-shu2, WU  Ke,  CONG Jian-bo,  XIAN Hong, WANG Chang-zheng, ZHANG Qing-jun   

  1. 1.Analysis and Measurement Center Jilin University;  2.Institute of Atom and Molecule Physics,Jilin  University;  3.Institute of Theoretical Chemistry,  Jilin University,Changchun 130023, China; Institute of Radiation Medicine, Academy of Military Medical Science, Beijing 100850, China
  • Received:2002-11-29 Revised:2003-03-03 Published:2003-06-05 Online:2003-06-05
  • Supported by:

    国家自然科学基金重点资助项目(29935080)

Abstract:

An apparatus for ESR and ESR imaging at L-band (1.05 GHz) was built in our laboratory, consisting of L-band ESR spectrometer, three sets of gradient coils and a personal computer-based data acquisition system. The sample cavity is a 3-loop 2-gap reentrant resonator, accepting samples with sizes up to 20 mm in diameter and 30 mm in length. The resonance frequency of unloaded resonator is 1.05 GHz. The microwave oscillator frequency is locked to the resonant frequency of loaded resonator by means of automatic frequency control (AFC). Linear magnetic field gradients (up to 2 mT/cm in amplitudes) in the X-, Y-, and Z-directions are produced by gradient coils, and have a linear range of 40 mm measuring from the center of the magnet. All projections in imaging were acquired as single scans with constant sweep time. To accurately compute the spatial distribution of the spins, the line shape function that causes line-broadening of the ESR signals needs to be removed from the observed projections. Three- dimensional ESR image is constructed based on the Lauterbur's method. Using this system, we studied the spatial distribution of TEMPO (2,2,6,6-tetramethylpiperidine-1-oxyl) nitroxide radicals in sample tube. Stack images of ESR-CT and 2D, 3D ESR images were obtained.

Key words: ESR, imaging system, L-band, field gradient, 3D image construction, ESR-CT