THE EFFECT OF FILLING FACTOR ON SENSITIVITY OF NMR SPECTROMETER

Li Jieshen, Wu gang   

  1. Beijing Analytical Instument Factory
  • Received:1984-09-15 Published:1985-03-05 Online:2018-01-23

Abstract: The effect of filling factor of receiver coil and the construction parameters of reciever coil on sensitivity of NMRspectrometer and The relationship between the limitation of Signal-Noise ratio for NMR signal and the tested Signal noice ratio for double T-bridge circuit have been discussed, An improvement has been taken to normal receiver coil (it is wound outside a tube which has 7mm O.D. and 6mm I.D.), The reciever coil is put inside the tube against its wall. It has been proved by experiment that using the CH203 NMR spectrometer (made by Beijing Analytical Instrument Factory) with 5mm O.D.tube and 1% ethyl benzene, The sensitivity is raised from 20:1 (by normal receiver coil) to 30:1 (by improved reveiver coil).