Chinese Journal of Magnetic Resonance ›› 1985, Vol. 2 ›› Issue (3): 249-252.

Previous Articles     Next Articles

DEMAGNETIZATION FACTOR, GEOMETRICAL FACTOR AND SUSCEPTIBILITY CORRECTION FORMULA FOR CYLINDRICAL SAMPLE UNDER UNIVERSAL CONDITIONS

Mao Xian, Shen Lianfang, Wu Qinyi   

  1. Wuhan Institute of Physics, Academia Sinica
  • Received:1984-12-10 Published:1985-09-05 Online:2018-01-23

Abstract: On the hypothesis of infinite length cylindrical medium,demagnetization factor, geometrical factor and susceptibility correction formula for arbitrary angular orientation of NMR sample tube with reference to the magnetic field are derived. It is concluded that under magic angle condition, solution susceptibility effect could be eliminated, which is similar to that of ellipsoid medium sample by Garroway. Exprimental results agree well with the theoretical deduction obtained.