Chinese Journal of Magnetic Resonance

Previous Articles     Next Articles

THE MEASURREMENT OF THE EFFECTIVE TRANSVERSE RELAXATION TIME IN THE PRESENCE OF RADIATION DAMPING

LUO Hui-jun   

  1. Analysis and Research Center, East China University of Science and Techn ology, Shanghai 200237
  • Received:2001-01-08 Published:2001-03-05 Online:2001-03-05

Abstract:

A new procedure for measuring the effective transverse relaxation time T*2 in the presence of radiation damping is proposed here. After the amplitude of the echo obtained by CPMG pulse sequence is largely decayed by the intrinsic T2 relaxation process, its linewidth is no longer broadened by the radiation da
mping effect and thus a good measure of the true value of T*2 is possible. Accordingly, for the damped case, the T*2 and T2 can in fact be simultaneously determined in a CPMG spin-echo experiment. In addition, when the saturation recovery experiment was utilized to measure T*2, the small-flip-angle pulse must be applied to detect the signals.

Key words:

very, NMR


CLC Number: