Abstract: The present way of microwave power adjustment of Bruker ESR spectrometer (ER-200 D; ESP-300) is digital. It provides a skipping power output by a DB appliance. According to its controlling principle, we realized to improve the way of microwave power adjustment to a continuous one without much more changing its original circuit. It is practicable and useful proved with some tests.
Jin An, Wu Ke. IMPROVEMENT OF MICROWAVE POWER ADJUSTMENT ON BRUKER ESR SPECTROMETER[J]. Chinese Journal of Magnetic Resonance, 1990, 7(2): 251-254.